Potential Safety Concern Involving Rosemount Pressure Transmitters

  • Type of Notification: Product Safety Notification
  • FM Approval Status: FM Approved

Alert Summary

FM Approvals has been notified by Emerson Process Management of a potential safety concern involving FM Approved Rosemount 2051T, 3051C, 3051D, 3051G, 3051T Pressure Transmitters with a display meter option (M4 – LCD Display with Local Operator Interface or M5 – LCD Display) manufactured from January 20 to January 22, 2015.

Company Identity

Emerson Process Management, Rosemount Tank Radar AB
Gamlestadsvägen 18 B, PO Box 13045
SE-402 51 GÖTEBORG
Sweden

Product Identity

Rosemount 2051T, 3051C, 3051D, 3051G, 3051T Pressure Transmitters shipped from January 20, 2015 and January 22, 2015.

Description

Emerson Process Management has identified a potential safety concern affecting the reliability of Rosemount models; 2051T, 3051C, 3051D, 3051G, 3051T with a display meter option code (M4 – LCD Display with Local Operator Interface or M5 – LCD Display) which were shipped from January 20, 2015 to January 22, 2015.

The potential safety concern was caused by an incorrect equipment setup in the manufacturing process for the meter cover assembly. The cemented joint between the metal and glass on the display meter cover may not comply with explosion proof requirements. Also, the cemented joint has the potential to allow environmental exposure to the transmitter electronics.

Emerson’s quality monitoring processes identified this issue. As a direct result of quality monitoring, production and shipping were immediately stopped as soon as the issue was identified. Formal corrective action has been taken to prevent the issue from re-occurring.

Make/Model

Rosemount 2051T, 3051C, 3051D, 3051G, 3051T Pressure Transmitters with a display meter option code (M4 – LCD Display with Local Operator Interface or M5 – LCD Display)

Nameplate Data

Rosemount 2051T, 3051C, 3051D, 3051G, 3051T Pressure Transmitters with a display meter option code (M4 – LCD Display with Local Operator Interface or M5 – LCD Display)

Hazard Involved

Emerson Process Management, Rosemount Inc has issued a product safety notification regarding FM Approved Rosemount 2051T, 3051C, 3051D, 3051G, 3051T Pressure Transmitters with a display meter option code (M4 – LCD Display with Local Operator Interface or M5 – LCD Display)manufactured from January 20 to January 22, 2015. 

Using This Product?

If you suspect that you are in possession of any of the affected equipment listed above bearing the FM Approvals certification marking, please bring it to the attention of: 

Antonio Pires

Norwood, MA, United States

+1 (1) 781-255-4825

Email